我们的服务

用户登陆

电子邮箱  
密码      忘记密码?
  注册
我们的服务
HPLC仪器确认指南
来源: | 作者:prodcc054 | 发布时间: 2017-08-17 | 664 次浏览 | 分享到:
 

 OMCL Network of the Council of Europe

QUALITY MANAGEMENT DOCUMENT

PA/PH/OMCL (11) 04

QUALIFICATION OF EQUIPMENT

仪器确认

ANNEX 1: QUALIFICATION OF HPLC EQUIPMENT

附件1HPLC仪器的确认

Full document title and reference

文件全名和索引号

Qualification of Equipment

Annex 1: Qualification of HPLC equipment

PA/PH/OMCL (11) 04

Document type 文件类型

Guideline 指南

Legislative basis 立法基础

The present document was also accepted by EA as recommendation document to be used in the context of Quality Management System audits of OMCLs

Date of first adoption

首次发行日期

May 2005

Date of original entry into force

首次执行日期

June 2005

Date of entry into force of revised document 修订后执行日期

1st July 2011

Previous titles/other references

原文件名/其它索引号

This document replaces document PA/PH/OMCL (07) 17 DEF

Custodian Organisation

保管机构

The present document was elaborated by the OMCL Network/ EDQM of the Council of Europe

Concerned Network

相关网络

GEON


ANNEX 1 OF THE OMCL NETWORK GUIDELINE

“QUALIFICATION OF EQUIPMENT”

OMCL网络指南“仪器的确认”之附件1

QUALIFICATION OF HPLC EQUIPMENT

HPLC仪器的确认

Introduction 概述

The present document is the first Annex of the core document “Qualification of Equipment”, and it should be used in combination with it when planning, performing and documenting the HPLC equipment qualification process.

本文件是核心文件“仪器的确认”第1个附件,在计划、实施和记录HPLC仪器的确认过程时,应将本文件与核心文件一起使用。

The core document contains the general introduction and the Level I and II of qualification, common to all type of instruments, and the present annex contains HPLC instrument-related recommendations on parameters to be checked and the corresponding typical acceptance Limits, as well as practical examples on the Methodology that can be used to carry out these checks.

核心文件包括了第一级和第二级确认的通用介绍,适用于所有类型的仪器,本附件包括了HPLC仪器相关的需要检查的参数和相应典型的可接受标准的推荐,以及可用于进行这此检查的实用方法学举例。

When qualifying HPLC equipment, it should be noted that it is acceptable to check at Level III and IV several of the mentioned parameters at the same time in a combined test procedure (e.g. “overall” system performance test giving information on peak area precision, retention time precision, gradient reproducibility, etc).

在进行HPLC仪器确认时,应注意也可以对第三级和第四级几个提到的参数采用联合检测程序同时进行检查(例如“全面”系统性能检查,同时给出峰面积精密度、保留时间精密度、梯度重复性等的信息)


TABLE III 表三

Level III. Periodic and motivated instrument checks

第三级 定期主动仪器检查

Examples of requirements for HPLC instruments and detectors

HPLC仪器和检测器要求举例

Instrument module

仪器模块

Parameter to be checked

需检查的参数

Typical tolerance Limits

典型允差限度

Solvent delivery system

溶剂传送系统

  Flow rate 流速

  Proportioning accuracy and precision (gradient test) 配比准确度和精密度(梯度测试)

  Proportioning ripple 配比波纹

●±5 %

●±2

●≤0.2 %

Injector

进样器

l  Volume precision 进样量精密度

l  Carry-over 残留

l  RSD1.0 %

l  see Annex I 见附件1

Autosampler

自动进样器

l  Thermostatting accuracy and precision 恒温准确度和精密度

l  ±3 °C

Oven or cooling device

柱温箱或冷却装置

l  Thermostatting accuracy

恒温准确度

l  ± 2 °C

UV/DAD detector

UV/DAD检测器

l  Linearity 线性

l  Wavelength accuracy 波长准确度

l   r2 0.999

l  ±2 nm

Fluorescence detector

荧光检测器

l  Wavelength accuracy excitation波长准确度激发

l  Wavelength accuracy emission 波长精密度发射

l  Sensitivity 灵敏度

l  ±3 nm

l  ±3 nm

l  ±see Annex I

Electrochemical detector 电化学检测器

l  Accuracy of the signal 信号精密度

l  Stability of the signal 信号稳定性

l  see Annex I 见附件1

l  see Annex I 见附件1

RID detector

RID检测器

l  Signal/Noise ratio 信噪比

l  Drift over time 飘移

l  see Annex I见附件1

l  0.1 mV/min

CD detector

CD检测器

l  Signal/Noise ratio 信噪比

l  Sensitivity 灵敏度

l  Drift over time 飘移

l  Linearity 线性

l  see Annex I见附件1

l  see Annex I见附件1

l  see Annex I见附件1

l  r > 0.999


TABLE IV 表四

Level IV. In-use instrument checks

第四级 在用仪器检查

Examples of requirements for HPLC instruments with UV or DAD detectors

配备UVDAD检测器的HPLC仪器的要求举例

Parameter to be checked

需检查参数

Typical tolerance Limit限度s

l  System suitability check for the Method方法系统适用性检查

According to Ph. Eur. or MAH dossier or validated in-house Method

根据欧洲药典或MAH文件或公司内部经过验证的方法

l   Peak area precision (applicable to the main peaks in the test solution)

峰面积准确度(适用于供试液中主峰面积)

RSD ≤ 1.5 %

(unless otherwise prescribed in the system suitability of the Method, e.g. specific requirements from Ph. Eur.2.2.46, API monographs or MA dossiers) 除非在方法的系统适用性里有描述,例如EP2.2.46特定要求,原料药各论或MA文件

l  Retention time precision

保留时间精密度

RSD ± 5 %

l  Carry-over (by comparing consecutive standard (of the substance being quantified) and blank injections) 残留(通过比较连续(进行定量检测的物质)标准和空白进样)

≤ 0.2 %

l  Signal/Noise ratio (to be applied for related substances test)

信噪比(用于有关物质检测)

According to Ph. Eur.

根据欧洲药典


ANNEX I 附件1

Level III. Periodic and motivated instrument checks

第三级 定期性的主动仪器检查

This Annex contains practical examples of tests and their associated tolerance Limits for several parameters related to the performance of the different modules of a HPLC.

本附件包括一些实用几个与HPLC不同模块相关的参数检测举例及其相应的可接受限度。

These examples can be considered by the OMCLs as possible approaches to perform the Level III of the equipment qualification process: “Periodic and motivated instrument checks”.

这些例子可以认为是OMCL实验室可能的进行第三级仪器确认的方法“周期主动仪器检查”

HPLC SOLVENT DELIVERY SYSTEM 高压液相溶剂传输系统

The following tests are proposed for the periodic and motivated check of the HPLC solvent delivery system: flow rate and gradient test.

以下测试作为HPLC溶剂传输系统周期性主动检查:流速和梯度检查

FLOW RATE 流速

Materials: 材料

Volumetric flask of 5 or 10 ml

5ml10ml的容量瓶

Calibrated chronometer

校正过的气压计

Settings: 设置

Mobile phase: degassed water

流动相:脱过气的水

No column (open end)*

无柱(开放式)*

Flow rate: adjusted between 0.5 and 3.0 ml/min

流速:调节在0.53.0ml/min

If high-pressure mixing systems are installed, this test has to be done on each solvent channel.

如果有安装高压混合系统,本测试应在每一溶剂通道进行。

* For certain equipment, e.g. in the case of low flow rates, the check would be performed by using a column or a backpressure regulator.

* 对于一定的仪器,例如,流速较低时,采用一根柱子或背压调整器进行检查

Method方法:

Set the flow rate at an appropriate level and measure the time needed to fill the volumetric flask up to the mark. Record the time needed.

将流速设定在一个合适的水平,测量将容量瓶装满至刻度的时间。记录所用的时间。

 f V /t  f (V×60)/t

f ________. measured flow rate [ml/min] 测量的流速

t ________. elapsed time to fill up to mark [s] 充满至刻度所花的时间

V _______. volume of the volumetric flask [ml] 容量瓶的体积

 D100×(fF)/F

D _______. deviation [%] 偏差

F _______. adjusted flow rate [ml/min] 调节后的流速

f ________. measured flow rate [ml/min] 测量的流速

Limit限度s: ±5%

GRADIENT COMPOSITION AND RIPPLE 梯度组成和波动

Settings: 设置

Stainless steel capillary e.g. 2000 x0.12 mminstalled instead of a column

不锈钢毛细管,例如2000X0.12mm,代替色谱柱安装于仪器

Detection: UV-Detector adjusted to 265 nm

检测器:紫外检测器调节至265nm

Mobile phase A: degassed water

流动相A:脱气水

Mobile phase B: degassed water containing 0.5% acetone

流动相B:含0.5%的脱气水

Flow rate: 1.0 ml/min

流速:1.0ml/分钟

Method方法:

The test is carried out in the following way by using a gradient program depending on the number of solvent channels and the configuration of the system:

按下列方法进行测试,根据通道数量不同使用下列不同系统参数和梯度进行洗脱

A-B

A-B and A-C

A-C, A-B and B-D

time [min]

时间【分钟】

% mobile phase A (water)

%流动相A(水相)

% mobile phase B

%流动相B

(water-acetone mixture)

(水-丙酮混合物)

0.0

100

0

0.1

90

10

10

90

10

10.1

50

50

20

50

50

20.1

10

90

30

10

90

30.1

0

100

40

0

100

40.1

100

0

Start the test by pumping water for at least 10 min to equilibrate the system.

开始测试前,先用水平衡系统至少10分钟。

The zero % value at the start of the test is the baseline. All steps are measured at the beginning of the horizontal part of the line either by software or manually on the paper print using a liner. The height of the 100% water/acetone mixture is used as the 100% value in the following calculation.

开始时的0%值是基线值。所有步骤均在开始时测量基线的水平部分,由软件或手动在纸上用衬垫打印。100%/丙酮混合物的高度在以下计算中作为100%值。

 %H100* h/H

%H ______ calculated composition 计算组成

h ________ height of the measured line 测量线的高度

H _______. height of the 100% water/acetone mixture line (mobile phase B)
100%
/丙酮混合物线的高度(流动相B

 d %HG

d ________ deviation 偏差

G _______. gradient composition adjusted [% acetone/water solution mixture = mobile phase B]
梯度组成调节【%丙酮/水溶液混合物 = 流动相B

Limit限度s:

Absolute deviation: ±2 of the adjusted value

绝对偏差: 调整值的±2

The ripple of the gradient composition is the percentage of noise of the 50% line from the gradient program.

梯度组分的波动为梯度程序噪声基线的50%

 %R 100* N /h50

%R ______ ripple 波纹

h50 ______. height of the 50% line 50%的线高度

N _______. height of the noise line, measured during 1 minute in the linear region 噪声线高度,在线性区域1分钟测量的结果

Limit限度s: ≤ 0.2 %

HPLC INJECTOR 进样器

Volume precision and carry-over are the tests proposed for the periodic and motivated check of the HPLC injector.

建议在HPLC进样器周期性主动检查中进行体积精密度和残留检查

VOLUME PRECISION AND CARRY-OVER

体积精密度和残留

Solutions: 溶液

Solvent A: methanol : water R, 60 : 40.

溶液 A:乙醇:水(试剂) = 6040

Reference solution (a): dissolve 15.0 mg methyl-, ethyl-, and propylparabene in 100.0 ml of solvent A.

对照液(a):取15.0mg乙基、甲基和尼泊金丙酯溶于100.0ml溶剂A

Reference solution (b): Dilute 1.0 ml of reference solution (a) to 10.0 ml of solvent A.

对照液(b):取1.0ml对照液(a)用溶于10.0ml溶剂A

Reference solution (c): Dilute 1.0 ml of reference solution (b) to 100.0 ml of solvent A.

对照液(c):取1.0ml对照液(b)用溶于100.0ml溶剂A

Settings: 设置

Column: Lichrospher 100 RP8, 5um, 125 x4 mm, without precolumn

色谱柱:Lichrospher 100 RP8, 5 um, 125 x4 mm, 无衬管

Mobile phase: methanol : water = 60 : 40

流动相:乙醇:水 = 6040

Flow rate: 1.0 ml/min

流动速度:1.0ml/分钟

Detection: 254 nm

检测器:254nm

Injection volume: 20 μl

进样体积:20ul

Method方法:

Injection scheme: 进样序列

      6x reference solution (b) 对照液(b6

     1x reference solution (a) 对照液(a1

      1x solvent A (blank injection 1) 溶剂A(空白)1

      1x reference solution (b) 对照液(b1

      1x solvent A (blank injection 2) 溶剂A(空白21

      1x reference solution (c) 对照液(c1

Limit限度s:

Repeatability of peak areas: The relative standard deviation of the peak areas of all peaks in the chromatogram obtained with the reference solution (b) should be 1.0 %.

峰面积重复性:对照液(b)色谱图中所有峰的峰面积相对标准偏差应1.0 %

Carry-over: The percentage of the peak area corresponding to propylparabene in the blank injection 1 does not exceed 0.5% of 10 times the peak area of the propylparabene peak in the chromatogram obtained with the reference solution (b) injected after the blank injection.

残留:尼泊金丙酯在空白图谱中的峰面积不超过在空白后进样的对照液(b)中尼泊金丙酯的对应峰面积的10倍的0.5%

The percentage of the peak area corresponding to propylparabene in the reference solution (c) is 0.9 – 1.1% of the peak area of the propylparabene peak in the chromatogram obtained with the reference solution (b) injected after the blank injection.

在对照液(c)中尼泊金丙酯对应的峰面积为空白后进样的对照液(b)中尼泊金丙酯峰面积的0.9-1.1%

HPLC AUTOSAMPLER 液相色谱仪自动进样器

Thermostatting accuracy and precision can be tested in the frame of the periodic and motivated check of the HPLC Autosampler.

HPLC自动进样器的周期主动检查可以进行温度准确度和精密度检查

THERMOSTATTING ACCURACY 温度准确性

Materials: 材料

Calibrated temperature probe.

校正过的温度探头

Method方法:

Select a temperature along the operational or required temperature range of the equipment. Wait until the system is equilibrated.

在可操作范围或要求的温度范围内选择并设定一个温度,等待系统平衡

By means of the calibrated probe, measure the actual temperature in the autosampler and compare it to the selected temperature.

采用校正过的温度探头,测量自动进样器的实际温度,与设定的温度进行比较

Repeat the same procedure at different pre-selected points covering the temperature range.

在选定的温度范围内不同温度点重复相同的步骤

Limit限度s:

The actual temperature may not differ more than ±3°C with respect to the selected temperature.

实际温度与设定的温度相比差值不超过±3°C

THERMOSTATTING PRECISION 温度准确性

Materials: 材料

Calibrated temperature probe.

校正过的温度探头。

Method方法:

Select a temperature along the operational or required temperature range of the equipment. Wait until the system is equilibrated.

选择一个操作温度或要求的仪器温度范围,等待系统平衡。

By means of the calibrated probe, make “n” measurements over a pre-established period of time. Compare the mean of the “n” measurements to the selected temperature.

采用校正过的温度探头,在预设的时间内进行n次测量,比较n次测量值的平均值与设定的温度。

Limits限度:

The actual temperature may not differ more than ±3°C with respect to the selected temperature.

实际温度与设定的温度之间差值不超过±3°C


HPLC OVEN/COOLING DEVICE  色谱仪柱温箱/降温设施

Thermostatting accuracy is the parameter tested in this example of periodic and motivated check of the HPLC oven/cooling device.

热力学准确性是在本例对HPLC柱温箱/降温设施周期主动检查的参数

THERMOSTATTING ACCURACY 温度准确性

Materials: 材料

Calibrated thermometer.

校正过的温度计

Method方法:

Set the column oven temperature to40 °C, wait about 30 minutes to equilibrate the system, put a calibrated thermometer into the oven and read the temperature after 10 minutes.

设定柱温箱温度为40 °C,等待约30分钟平衡系统,将校正过的温度计放入柱温箱中,在10分钟后读取温度。

Limits限度: 38 -42°C.

HPLC UV/DAD DETECTOR 紫外/DAD检测器

The periodic and motivated check of the HPLC UV/DAD detector can be performed by testing the linearity and the wavelength accuracy.

对于HPLC紫外/DAD检测器的周期主动检查,可以对线性和波长准确性进行检查

LINEARITY 线性

Solutions: 溶剂配制

Std. 1: 0.5 μg caffeine/1 ml methanol HPLC Grade

标准2 0.5 μg咖啡因/1ml色谱级甲醇

Std. 2: 1.0 μg caffeine/1 ml methanol HPLC Grade

标准2 1.0 μg咖啡因/1ml色谱级甲醇

Std. 3: 5.0 μg caffeine/1 ml methanol HPLC Grade

标准3 5.0 μg咖啡因/1ml色谱级甲醇

Std. 4: 25.0 μg caffeine/1 ml methanol HPLC Grade

标准425.0 μg咖啡因/1ml色谱级甲醇

Std. 5: 50.0 μg caffeine/1 ml methanol HPLC Grade

标准550.0 μg咖啡因/1ml色谱级甲醇

Std. 6: methanol HPLC Grade (blank)

标准6:色谱级甲醇(空白)

Std. 5: weigh 9.0 to 11.0 mg caffeine and fill up to 200.0 ml with methanol HPLC

标准5:称取9.011.0mg咖啡因,用HPLC级别甲醇稀释至200.0ml

Std. 4: dilute 50.0 ml of Std. 5 to 100.0 ml with methanol

标准4:取50.0ml标准5溶液用甲醇稀释至100.0ml

Std. 3: dilute 10.0 ml of Std. 5 to 100.0 ml with methanol

标准3:取10.0ml标准溶液5用甲醇稀释至100.0ml

Std. 2: dilute 20.0 ml of Std. 3 to 100.0 ml with methanol

标准2:取20.0ml标准3溶液用甲醇稀释至100.0ml

Std. 1: dilute 10.0 ml of Std. 3 to 100.0 ml with methanol

标准1:取10.0ml标准3用甲醇稀释至100.0ml

Settings: 设置

Column: RP-18 5 um 30-50 x 2,1-4.6 mmor capillary2000 mmx0.12 mmID

色谱柱:RP-185 um 30-50 x 2,1-4.6 mmor capillary2000 mmx0.12 mmID

Mobile phase: methanol HPLC Grade

流动相:色谱级甲醇

Oven temperature:40 °C

柱温箱温度:40 °C

Flow rate: 1.0 ml/min (adjusted by using 100% methanol)

流速:1.0ml/min(用100%甲醇调整)

Detection: 273 nm

检测波长:273nm

Injection volume: 20μl

进样量:20ul

Method方法:

Injection scheme:

进样序列

2 x blank

1 x Std. 1

1 x Std. 2

1 x Std. 3

1 x Std. 4

1 x Std. 5

2针空白,标准1-5依序各一针

Limits限度: r2 ≥ 0.999

Remark: As this test employs different test solutions to be injected, it covers also the check of correct positioning vials in the autosampler.

注:由于此检测中要注入不同溶液,因此也包括了自动进样器样品瓶位置正确性的确认

WAVELENGTH ACCURACY 波长准确度

If there are built-in test procedures for the determination and adjustment of wavelength accuracy, follow the instructions of the instrument manual.

如果内置检测程序可以检测和调节波长准确度,则根据操作手册进行操作

In all other cases use the procedure described below.

在所有其它情况下采用以下程序

Solutions: 溶液

DAD: caffeine Std. 5 from the linearity testing 线性测试中咖啡因标准溶液5

UV/VIS: 1.0 μg/ml anthracene solution in water 蒽水溶液

Settings: 设置

Mobile phase: 15% acetonitrile in water

流动相:15%乙腈水溶液

Column: RP18, 5 m 30-50 x 2,1-4.6 mmor capillary2.0 mx0.12 mmID

色谱柱:RP185μm, 30-50×2.1-4.6mm或毛细管柱20m×0.12mm内径

Oven temperature:40 °C

柱箱温度:40

Flow rate: 1.0 ml/min (adjusted by using 15% acetonitrile in water)

流速:1.0ml/min(用15%的乙腈水溶液调节)

Detection: scan from 230 nm to 290 nm (DAD)

检测:自230290nm扫描(DAD

Injection volume: 20 μl

进样体积:20μl

Method方法:

DAD: Inject 20μl of the caffeine solution and record the spectrum. The maximum is at 272 nm and the minimum at 244 nm.

DAD: 咖啡因溶液进样20μl,记录色谱图,最大吸收波长为272nm,最小为244nm.

UV/VIS: fill the cell with the anthracene solution and change the wavelength from 248 to 254 nm in 1 nm steps; record the maximum of absorption. The theoretical value is 251 nm.

UV/可见光:向池中注入蒽溶液,将波长以1nm的步距从248变化至254nm,记录最大吸收值。理论值为251nm

Limit限度s: 2 nm

HPLC FLUORESCENCE DETECTOR 荧光检测器

The following three parameters are proposed for the performance of the periodic and motivated check of the HPLC fluorescent detector:

下面三个参数建议作为荧光检测器性能周期自主检查

WAVELENGTH ACCURACY EXCITATION 波长激发准确度

Method方法: 方法

Rinse and fill the measuring cell with de-ionized water

用去离子水淋洗并充满检测池

Adjust the excitation wavelength to 350 nm.

调节激发波长为350nm

Measure the emission and subtract 397 nm (theoretical value).

测量发射波长减去397nm(理论值)

Limit限度限度s: ±3 nm

WAVELENGTH ACCURACY EMISSION 波长发射准确度

Method方法: 方法

Rinse and fill the measuring cell with de-ionized water

用去离子水淋洗并充满检测池

Adjust the emission wavelength to 397 nm.

调节发射波长为397nm

Measure the excitation and subtract 350 nm (theoretical value).

测量激发波长减去350nm(理论值)

Limit限度限度s:±3nm

SENSITIVITY 灵敏度

Solutions: 溶液

Quinine HCl.2H2O solution conc. 0.015μg/ml (=15 ppb)

二水盐酸喹啉溶液,浓度0.015μg/ml (=15 ppb)

The quinine solution is prepared with the following mobile phase: dissolve6.8 gof potassium dihydrogen phosphate R and3.0 gof hexylamine R in 700 ml of water R, adjust to pH 2.8 with dilute phosphoric acid R, add 90 ml of acetonitrile R and dilute to 1000.0 ml with water R.

喹啉溶液用以下流动相制备:溶解6.8g磷酸二氢钾和3.0g己胺于700ml试剂级水中,用稀磷酸调pH2.8,加90ml乙腈并用试剂级水稀释至1000.0ml.

Settings: 设置

The chromatographic conditions are set according to Ph. Eur. “Quinine HCl” (01/2005:0018), test “Other cinchona alkaloids”, with modified flow rate and acetonitrile concentration. 色谱条件设定依据欧洲药典“盐酸喹啉”(01/2005:0018),检测“金鸡纳生物碱”,改变流速和乙腈浓度。

Mobile phase: as above

流动相:同上

Column: RP18, 5 μm, 250 × 4.6 mm

色谱柱: RP18, 5 μm, 250 ×4.6 mm

Flow rate: 1.2 ml/min

流速:1.2 ml/min

Excitation wavelength: 350 nm

激发波长:350nm

Emission wavelength: 397 nm

发射波长:397nm

Flow-cell volume: 8 μl (for this example, a Waters 2475 Multi Fluorescentie detector was used. Flow-cell volume may vary depending on the instrument manufacturer)

流动池体积:8μl(此例中,使用的是Waters2475多荧光检测器,流动池体积可能根据仪器厂商不同而不同)

Method方法:

Inject 10μl of the quinine solution and measure the peak height.

进样喹啉溶液10μl,测量峰高度

Inject 10μl of the blank and measure the peak height of the noise.

进样空白溶液10μl,测量噪音峰高

Divide the peak height of the quinine solution by 3 times the peak height of the noise.

将喹啉溶液峰高除以3倍噪音峰高

Divide the concentration of the quinine solution by the previously obtained factor.

将喹啉溶液浓度除以之前的系数

Limit限度限度s: ≤ 0.5 ppb

HPLC ELECTROCHEMICAL DETECTOR 高效液相电化学检测器

Accuracy and stability of the signal are the proposed parameters to be tested during the periodic and motivated check of the HPLC electrochemical detector.

电化学检测器的信号准确度和稳定性是定期自主检测中需要测定的参数

ACCURACY AND STABILITY OF THE SIGNAL 信号准确度和稳定性

Settings: 设定

Cell potential of a dummy cell: 800 mV

空白池电压:800mv

Rise time filter: 0.1 s

增加时间过滤:0.1s

Range: 0.1 nA

范围:0.1nA

Temperature:30 °C

温度:30

Method方法:

Accuracy: Measure the electric current and subtract 2.67 nA (theoretical value)

准确度:检测电流并减去2.67nA(理论值)

Stability: Measure the noise over a period of 5 minutes

稳定性:连续5分钟测量噪音

Limit限度s:

Accuracy (cell current):  0.05 nA

准确度(池电流):0.05nA

Stability of the signal (Noise): max. 2 pA or 20 mV

信号(噪音)稳定性:最大2pA20mV

HPLC RID DETECTOR 高效液相RID检测器

Signal to Noise ratio and drift over time are the parameters proposed for the periodic and motivated check of the HPLC RID (refractive index) detector. 高效液相RID(折光率)检测器的信噪比和对时间的漂移是定期自主检测中需要测定的参数

SIGNAL TO NOISE RATIO 信噪比

Solutions溶液:

Standard solution: D-fructose concentrate solution at 4.0 mg/ml (dilute 200.0 mg fructose + 20 ml water + 25.0 ml acetonitrile up to 50.0 ml with water for HPLC)

标准溶液:D-左旋糖溶液4.0mg/ml(将200.0mg左旋糖+20ml+25.0ml乙腈,用色谱纯水稀释至50.0ml

Settings设置:

Column: spherisorb NH2 (or equivalent) 250 ×4.6 mm or other

色谱柱:spherisorb NH2 (或相当的) 250 ×4.6 mm 或其他

Oven temperature:38°C

柱温箱温度:38

Flow rate: 1.0 ml/min

流速: 1.0 ml/min

Injection volume: 20 μl

进样量: 20 μl

Mobile phase:0.253gsodium hydrogen phosphate R in 220 ml + 780 ml acetonitrile

流动相:0.253g磷酸氢二钠溶于220ml+780ml乙腈

Method方法:

After equilibration, inject three times a blank solution of mobile phase over a run time where the system is stable. Measure the baseline noise over an appropriate period.

平衡后,以流动相为空白进样三针,系统稳定运行适当时间,测定适当时长内噪音基线

The baseline noise is accepted if the mean height of the three replicates is < 1000 μV.

可接受基线噪音:三次平行进样平均值< 1000 μV

To calculate the signal to noise ratio, inject three times a solution of fructose at 0.4 mg/ml and calculate the mean of the three replicates.

计算信噪比,左旋糖溶液0.4mg/ml进样三针,计算平均值

Limit限度s: S/N ≥10

限度:信噪比≥10

DRIFT OVER A DEFINED PERIOD OF TIME 指定时长漂移

Method方法: 方法

Calculate the slope of the amplitude of random variations in the detector’s signal over 1 minute.

计算任意一分钟信号随机波动幅度的斜率

Limit限度s: ±0.1 mV/min

限度:±0.1 mV/min

Alternatively, the requirement may be expressed in RI/min or in % of full scale of the selected range.

也可以表达为RI/min或者以选择量程的全范围%

HPLC CD DETECTOR 高效液相色谱CD检测器

The following tests are proposed to perform the periodic and motivated check of the HPLC CD (circular dichroism) detector.

建议配备有CD(圆二图谱)检测器的HPLC进行以下周期和主动检查

LINEARITY AND SIGNAL TO NOISE RATIO

线性和信噪比

Solutions: 溶液

Reference solution (a): dissolve 25.0 mg D(-) pantolactone in 50.0 ml water

对照溶液(a):溶解25.0mg D-)泛酰内酯于50.0ml水中

Reference solution (b): dilute 2.0 ml of reference solution (a) to 10.0 ml with water

对照溶液(b):取对照溶液(a2.0ml用水稀释至10.0ml

Reference solution (c): dilute 4.0 ml of reference solution (a) to 10.0 ml with water

对照溶液(c):取对照溶液(a4.0ml用水稀释至10.0ml.

Reference solution (d): dilute 6.0 ml of reference solution (a) to 10.0 ml with water

对照溶液(d):取对照溶液(a6.0ml用水稀释至10.0ml.

Reference solution (e): dilute 8.0 ml of reference solution (a) to 10.0 ml with water

对照溶液(e):取对照溶液(a8.0ml用水稀释至10.0ml.

Reference solution (f): dilute 0.5 ml of reference solution (b) to 25.0 ml with water

对照溶液(f):取对照溶液(b0.5ml用水稀释至25.0ml.

Settings: 设定

Column色谱柱: C18, 150 ×4 mm, 5 μm

Mobile phase流动相: acetonitrile乙腈 : water = 10 : 90

Flow流速: 1.0 ml/min

Detection检测波长: 225 nm

Injection volume进样量: 20 μl

Method方法:

Check the linearity of de CD- and UV-signal of D(-)pantolactone reference solution a,b,c,d,e.

检查CDUVD-)泛酰内酯对照溶液a,b,c,d,e信号的线性关系

Measure the noise of the CD-signal of reference solution (f) between 0 – 10 min.

检查CD对对照溶液(f)在1-10分钟的噪声

- Calculate the absolute concentration (μg) in the cell

 计算池中绝对浓度(μg)

- Calculate the signal-to-noise ratio (S/N) for 0.01 μg in the cell

 计算池中0.01 μg信噪比(S/N

-          Calculate the sensitivity with the calculated S/N and the specified S/N= 2 (0.01 x 2/ S/N calculated)

-          计算所得信噪比及信噪比为2(0.01 x 2/ S/N calculated)时的灵敏度

Limit限度s:

Linearity: The linearity of the calibration line obtained with reference solution (a,b,c,d,e) should be r > 0.999.

线性:由对照液(a,b,c,d,e)所得的校正曲线线性系统应> 0.999

Sensitivity: The sensitivity at S/N= 2 should be better than 0.01μg.

灵敏度:信噪比为2时灵敏度应好于0.01μg

S/N ratio: The Limit for S/N is > 1.0 and the sensitivity should be 0.020 μg at maximum.

信噪比:信噪比为1.0时的限度和灵敏度最大为0.020μg

DRIFT OVER A DEFINED PERIOD OF TIME 指定时间内飘移值

Settings设定:

Column色谱柱: C18, 150 x4 mm, 5μm

Mobile phase流动相: acetonitrile乙腈 : water = 10 : 90

Flow流速: 1.0 ml/min

Detection检测波长: 290 nm

Injection volume进样量: 20μl

Method方法:

Inject water and stop the flow after 5 minutes. Measure the CD-signal for 1 hour.

进水样,停泵,5分钟后测量CD信号,持续1小时。

Measure with the cursor the drift of the baseline between 5 and 65 min.

用标尺测量565分钟间的基线飘移

Limits限度:

Not more than 0.1 mdeg/h

不超过0.1 mdeg/h

SPECTRA COMPARISON 谱图比较

Solutions溶液:

Reference solution (a): dissolve 5.0 mg dexamethasone in 10.0 ml 40% acetonitrile

对照溶液(a):溶解5.0mg 地塞米松于10.0ml 40%乙腈中。

Settings设定:

Column色谱柱: C18, 150 x4 mm, 5μm

Mobile phase流动相: acetonitrile乙腈 : water = 40 : 60

Flow流速: 1.0 ml/min

Detection检测波长: 230 nm

Injection volume进样量: 20ul

Method方法:

Compare the maxima/minima obtained at the Installation of the detector (see table).

比较检测器安装结果(见下表)

CD max 最大

CD min 最小

UV max 最大

222 nm

224 nm

236 nm

230 nm

252 nm

284 nm

Limits限度:

The maxima and minima may not differ more than± 4 nm.

最大和最小之间差异不可超过4nm